Height with machine lid
Height without machine lid
Width
Depth
Depth with door open
Wash cabinet dimensions:
Height
Width
Depth of upper basket/lower basket
Weight (net)
Max. load capacity of open door
Voltage, rated load, fuse rating
Connection cable
Water temperature water connection:
Cold water/steam condenser
Hot water/demineralised water (optional)
Static water pressure
Minimum water connection flow pressure:
Cold water/steam condenser
Hot water
Demineralised water (optional)
Recommended water connection flow pressure:
Cold water/hot water
Demineralised water (optional)
Steam condenser
DI water connection without pressure (optional)
Delivery head
Drainage length
Operation (according to IEC/EN 61010-1):
Ambient temperature
Relative humidity maximum
linear decrease to
Relative humidity minimum
Storage and transportation conditions:
Ambient temperature
Relative humidity
Air pressure
Altitude above sea level (according to IEC/EN 61010-1)
Protection category (according to IEC 60529)
Soiling level (according to IEC/EN 61010-1)
Overvoltage category (according to IEC 60664)
Noise level in dB (A),
sound pressure LpA during cleaning and drying phases
Test certifications
marking
Manufacturer address
1,500
* If installed above
m, the boiling point of the wash water will be lower. In this case, the disinfecting temperature and the
holding time might need to be adjusted.
835
mm
820
mm
898
mm
598
mm
1,200
mm
520
mm
530
mm
474
520
mm/
mm
98
kg
37
kg
See data plate
Approx.
1.8
m
20
Max.
°C
65
Max.
°C
1,000
Max.
kPa pressure
100
kPa pressure
40
kPa pressure
30
kPa pressure
≥
200
kPa pressure
≥
200
kPa pressure
≥
100
kPa pressure
8.5-60
kPa
Min.
0.3
m, max.
1.0
m
Max.
4.0
m
5
40
°C to
°C
80 % for temperatures up to 31 °C
50 % for temperatures up to 40 °C
10 %
- 20
60
°C to
°C
10
85
% to
%
500
1060
hPa to
hPa
2,000
Up to
m *
20
IP
2
II
<
70
VDE, EMC radio suppressed
Machinery Directive 2006/42/EC
Miele & Cie. KG, Carl-Miele-Straße 29,
33332 Gütersloh, Germany
Technical data
179