Page 1
Operating and Installation Instructions Laboratory Glassware Washer PG 8527 / PG 8528 en - US, CA To prevent accidents and machine damage read these instructions before installation or use. M.-Nr. 06 749 640...
A damaged machine is dangerous. service agency in strict accordance Turn off the machine immediately at the with national and local safety power switch and call the Miele Service regulations and standards. Department. Personnel operating the machine should be trained regularly. Children...
Page 5
IMPORTANT SAFETY INSTRUCTIONS Take care when handling liquids Always allow mobile units, modules, such as detergents, neutralizing inserts, and loads to cool before agents, wetting agents and rinse aids. removing. Be sure to empty any water These may contain irritant or corrosive from concave items into the wash ingredients, acids or alkalis.
Page 6
Do not allow any remains of acids, by Miele. Suds can have an adverse solvents or corroding ferrous material, effect on the operation of the machine. and in particular hydrochloric acid or...
Page 7
Make the door lock inoperative or remove the door Only specific additional equipment completely, so that children cannot made by Miele should be connected to accidentally shut themselves in. Make this machine. Consult a Miele appropriate arrangements for the safe Application Specialist on the type and disposal of the machine.
Description of the machine The Miele PG 8527/ 8528 Laboratory In this operating manual, the general Glassware Washer is designed for terms labware and glassware are used high-throughput, centralized wash to cover a range of laboratory areas. This unit is ideal for cleaning equipment and medical items.
Page 9
Description of the machine Loading side a Housing unit for Drying System DS d Main Switch "On/Off" and/or Steam Condensor SC e Barcode scanner mount (optional) b "Profitronic" electronic control system f Service panel (see "Programming manual") g Docking station for transfer cart c Multiport MF 27/28 h Lift door (closed)
Page 10
Description of the machine Loading side a Lift door (open) e Container for dispensing systems DOS 1 - DOS 4 b Connection for transfer cart and modules c Lower spray arm d Filter combination...
Page 11
Description of the machine Unloading side (PG 8528 only) a Lift door (closed) e Printer (optional) b Service panel with docking station f "Door" button for transfer cart MF 27/28 g Display c Multiport d Barcode scanner mount (optional)
Page 13
-Graph during program number. See the "Programming manual" operation. for more information. d Optical Interface # used by Miele Service Technicians e Start button h start a program f Clear button +C – return to the previous level – discard entry...
Context menu Certain functions are available through the context menu at any time, regardless of the operating level: – change display language – change operating level, – enter Delay Start – for convertible washers: change heating type, – show current fault(s), (if applicable). ^ To enter the context menu, press and hold the buttons - e at the same time.
Context menu Change the display language You can use the context menu to change the language used in the display. This setting is stored until the washer is switched off with the s button or the main switch. The language set under Settings / Language J does not change.
Page 17
To change the codes for existing users, or to enter new users, you must be a registered administrator. The administrator can be registered by Miele. ^ Press and hold the - e buttons for at least 2 seconds. The context menu appears in the display:...
Page 18
Context menu ^ Select the menu item Change operating level , confirm with OK. Change operating level Operating level D Operating level A Operating level B ^ Select the desired level and confirm with OK. Select the user User ABC Administrator Technical Service ^ Select the user group and confirm with OK.
Context menu Delay start The start time for a program can be pre-selected. ^ Select Delay start and confirm. ^ To activate the Start Delay option, go to the menu item Activate and select Delay start Activate Delay start ^ To use this feature, select Delay start and confirm.
Page 20
Context menu The selected program, the delay start time, the current time of day, and the time remaining until the program starts are displayed. LAB-STANDARD Start time 5:15 3:20 1:55 Remaining time The selected program automatically starts at the delay start time.
Context menu Change the heating type Current faults In convertible labwashers, you can use This menu item can be used to read the the context menu to change the heating current fault messages. type - e. g. from electric to steam ^ Select the menu item and confirm heating.
Automatic mobile unit recognition (AWK) Mobile unit coding In operating level C the AWK (optional) attributes a fixed program place to a The automatic mobile unit recognition mobile unit with valid coding. feature assigns a program place to a Program places 1-15 are reserved for mobile unit.
Automatic mobile unit recognition (AWK) Mobile unit coding strip Setting mobile unit coding 15 different codes can be set. They are To set or alter the coding of a mobile assigned to program places 1 to 15. unit with automatic recognition (AWK) proceed as follows: Under "System function - Selector switch organization"...
Page 24
Mobile units without side coupling must be coded with a track that does not have Bit 6. The magnetic strips on the PG 8527 / PG 8527 must be black. ^ Put the magnetic strip back in the track. Important:...
– Are the detergent and neutralizer containers full? Check after cleaning – Check the cleaning results visually. – Perform protein analytical checks, e.g. with the Miele test kit. – Are the jets and connections securely connected with the mobile unit / inserts?
Application Loading tips ^ Tall, narrow pieces should be placed in the center of the baskets. This Select mobile units, baskets and ensures good water coverage. inserts appropriate for the ^ Mobile units with an adapter must application. engage correctly. ^ Load the items to be washed so that ^ Engage inserts correctly into the water can access all surfaces.
Application Laboratory glassware (LG) Loading the machine Wide-necked glassware, such as wide Removing excess soils necked erlenmeyer flasks and petri ^ Empty all glassware before loading dishes or glassware with a cylindrical form, e.g. test tubes, can be cleaned into the machine. Take any hygienic inside and out with the rotating spray measures necessary to avoid arm.
Adding liquid detergents and neutralizers The washer comes equipped, as Use only detergents and standard, with two internal dispensing neutralizers specially designed for systems: labwashers. Always observe the – Dispensing system DOS 1 for use manufacturer's recommendations for with alkaline products, e. g., liquid use.
Adding liquid detergents and neutralizers Dispensing systems Adding liquid detergents and neutralizers Four 10 l containers of processing chemicals can be placed behind the service panel in the washer. Any additional containers will need to be set up outside the washer. Use caution when handling processing chemicals.
Page 30
Adding liquid detergents and neutralizers Message "Check dispensing system [X]" The currently running program pauses. ^ Check the container(s) and dispensing hoses indicated in the display. Refill or replace as needed. Containers and dispensing systems that are not in use can be shut off, to avoid fault messages in the display (see "Machine function / Query container"...
Operation Main Switch - On/Off Switching On ^ Press and hold the s button . The main switch disconnects the labwasher from power supply The activating time for the s button ^ Turn the main switch to I-ON. can be set to a maximum of 10 Once the start procedure is completed, seconds (see "System functions the labwasher is ready to for use.
Opening and closing the door Starting a program ^ Press the a button. You can find detailed and important information regarding the Miele The door opens or closes. default programs in the "Program To open the door on the clean side chart"...
Page 33
Operation Operating levels A and B Operating level C A list of all selectable programs Make sure that no metallic parts appears in the display. or pieces are stuck to the magnetic ^ Select the desired program with - e, strip or its underside.
Operation Program sequence Canceling a program Once started, the program proceeds A program can only be interrupted automatically. The program steps are or canceled in operating levels B shown in the display. and D. Detailed information on the program ^ Press the +C button. sequences is available in the The program is interrupted.
Operation Interrupting a program Note PG 8528: Which door can be opened depends A program can only be interrupted on which setting has been programmed or cancelled in operating levels B under "Machine Function/Door and D. Mechanism" (see the Programming Manual).
Barcode scanner (optional) Barcode scanners can be connected to the washer through the multiport on the unclean side and, if available, on the clean side as well. The following functions are available for barcode scanning: – User scan load – User scan unload –...
The interface cable is connected to ready to send the Ethernet interface in the washer. 1-4-6-9 empty Only a Miele Technician can The interface configuration is connect it to a different interface. described in the section "PC/Print The supplied cable can be used to functions"...
Maintenance Labwashers should undergo routine Also during maintenance, the following maintenance by the Miele Technical function checks are conducted: Service every 1000 operating hours – a test program is run or at least once every 6 months. – wash temperatures are measured Maintenance covers the following: –...
Maintenance Process validation Routine checks As a rule, it is the responsibility of the The user must complete these checks user to ensure that items cleaned in the every day before using the washer. A labwasher meet the required checklist is supplied with the washer. standards.
Maintenance Cleaning the filters in the wash Cleaning the fine filter cabinet The filters in the base of the wash cabinet prevent large particles from entering the system. The filters can become clogged, they need to be checked regularly and cleaned as needed.
Maintenance Cleaning the filter system for the circulation pump To protect the circulation pump, there are two additional filters under the flat filter. ^ Pull the filters up and out of the holder to remove. Clean if necessary. ^ Replace the filter combination in reverse order.
If you notice wear and tear on the then rinse throughly under running bearings or the gasket, please water. contact Miele. To do so, remove the spray arms as ^ After cleaning, put the spray arms follows: back in place and tighten the nut.
Cleaning the exterior Damaged or leaky door seals should be replaced by Miele Technical ^ To clean the stainless steel exterior, Service. use a damp cloth and dish soap, or a non-abrasive stainless steel cleaner.
Maintenance Baskets and inserts If there are spray arms: To ensure that the mobile units, – Do the spray arms rotate freely? baskets, and inserts function properly, – Are the spray jets clogged? See the they must be checked routinely. A section on "Cleaning and Care / checklist is supplied with the Cleaning the spray arms".
Miele. re-emerges above the ribbon cartridge. ^ Guide the paper through the slit in the front panel. Shut the panel. Replacement paper rolls (2 1/4" (58mm) wide / outer diameter approx. 2" (50mm) can be obtained from Miele.
After sales service Should you have any difficulties please contact the Miele Technical Service Department. See back cover for contact information. ^ Please quote the model and serial number of your machine. This information can be found on the data plate (see also "Electrical...
Electrical connection This unit is convertible for use with All electrical work must be carried steam or electric heating. out by a qualified electrician in Hook up requirements are as follows: accordance with local and national safety regulations. 1. Electric heat only: 208 V 3 phase 60 HZ 60 amp –...
Plumbing Water connection The machine cannot be operated with an inflow pressure of less than 10 psi. This machine must be connected to The machine is supplied with the water supply in accordance with 3 connection hoses 5 ft. (1.5 m) long all national and local plumbing with "...
Page 50
A feed pump must be installed by Miele's Technical Service Dept. The discharge nipple of the DI water container (pressure less) must be at...
Technical data Height 98 7/16" 250 cm (minimum room height) Width 45 1/4" 115 cm Depth 34 1/4" 87 cm Weight (net) approx. 1210 lbs 550 kg Operating weight approx. 1760 lbs 800 kg Voltage (volts) see data plate Connected load (watts) see data plate Current (amps) see data plate...
Caring for the environment Disposal of the packing Disposal of an old appliance materials Old appliances may contain materials that can be recycled. Please contact The cardboard box and packing your local authorities about recycling in materials protect the appliance during your area.
Page 56
Subject to change without notice / 4007 M.-Nr. 06 749 640 / 00...